ADVIA Centaur Cyclosporine
Recall: Z-1687-2018
·
Initiated January 2, 2018
Recall
- Recall Number
- Z-1687-2018
- Event Number
- 79505
- Firm
- Siemens Healthcare Diagnostics, Inc.
- FEI Number
- 2432235
- Product Code
- MKW
- Status
- Terminated
- Root Cause
- Error in labeling
- Initiated
- January 2, 2018
- Posted
- May 4, 2018
- Terminated
- February 10, 2023
- Address
- 511 Benedict Ave, Tarrytown, NY, 10591-5005
Description
ADVIA Centaur Cyclosporine
Reason
The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere
Action
Siemens Healthcare Diagnostics issued a recall in March 2017 for products with incorrect units for biotin in the Instructions for Use. After further investigation, recalling firm determined that more Siemens products also had missing or incorrect information in the IFUs for which two recalls were initiated, and Amended Corrections and Removals Reports were submitted to the FDA
Distribution
Nationwide and Canada, Mexico, and Thailand