MEGA 8Fr 50cc Intra-Aortic Balloon Catheter

Recall: Z-1685-2022 · Initiated August 5, 2022

Recall

Recall Number
Z-1685-2022
Event Number
90684
Firm
Datascope Corporation
FEI Number
1000222374
Product Code
DSP
Status
Open, Classified
Root Cause
Vendor change control
Initiated
August 5, 2022
Posted
September 8, 2022
Address
15 Law Dr, Fairfield, NJ, 07004-3206

Description

MEGA 8Fr 50cc Intra-Aortic Balloon Catheter

Reason

Certain lots containing undersized dilator.

Action

The firm distributed an Urgent Medical Device Removal notification, by mail, dated August 5, 2022. The notification informs customers that certain lots of 8Fr. 50cc IAB catheter and insertion kits are being recalled due to an undersized vessel dilator. During a sheathless insertion, the undersized inner diameter may result in the inability to insert the guidewire through the inner lumen of the dilator and the undersized outer diameter may reduce dilation of the vessel. If used in a procedure, recalled product could result in a procedural delay, gas emoli, and vessel injury. Customers are asked to examine their inventory to determine if they have nay recalled product and to quarantine any affected units. Customers with recalled product are to contact Datascope Corp./Getinge Customer Service at 888-9GETUSA (888-943-8872, option 2) between the hours of 8 AM and 6 PM to request a return authorization number to receive full credit. Distributors who have shipped any product are to forward the notification to customers. The recall notification is to be forwarded to all users at hospitals/facilities. Customers are asked to complete and return the recall response form by email to [email protected] or by fax to 877-647-1921.

Distribution

Awaiting domestic distribution. Foreign: Australia, Belgium, Canada, Czech Republic, France, Germany, Hong Kong, Iceland, Ireland, Italy, Netherlands, New Zealand, Norway, Poland, Portugal, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.

Quantity

10,427 units