FDA Recall Open, Classified

HistoCore PELORIS 3, Model/Catalog Number: 45.0005/45.7512.501 A11, Software Version: 3.4.0. The HistoCore PELORIS 3 Rapid Tissue Processor is a dual retort rapid tissue processor used to prepare tissue samples.

Recall: Z-1683-2025 · Initiated April 2, 2025

Recall

Recall Number
Z-1683-2025
Event Number
96612
Firm
Leica Biosystems Melbourne Pty Ltd
FEI Number
3006461677
Product Code
IEO
Status
Open, Classified
Root Cause
Release of Material/Component prior to receiving test results
Initiated
April 2, 2025
Posted
April 28, 2025
Address
495 Blackburn Road, Mount Waverley, Australia

Description

HistoCore PELORIS 3, Model/Catalog Number: 45.0005/45.7512.501 A11, Software Version: 3.4.0. The HistoCore PELORIS 3 Rapid Tissue Processor is a dual retort rapid tissue processor used to prepare tissue samples.

Reason

There is a leakage issue associated with the tubing in the manifold of the instrument.

Action

An URGENT MEDICAL DEVICE RECALL notification letter dated 3/31/25 was sent to customers. ACTIONS REQUIRED 1. If leakage from the instrument is observed in the drip tray or surrounding the instrument, please follow your laboratory's leakage containment/safety protocol. If a protocol is running, wait for the run to finish, switch off the instrument and contact your local Leica Biosystems representative immediately. 2. If no leakage has been observed from the instrument, continue using it as normal. Conduct routine monitoring for leakage until the on-site inspection is completed. 3. Your local Leica Biosystems representative will contact you to schedule an onsite inspection of your instruments tubing and reseat the tubes if needed. 4. Confirm Receipt of Recall Notice Please confirm receipt of this notice by signing and dating the attached Recall Acknowledgement Form by email to the following email address: lbsme/[email protected] Please contact your local Leica Biosystems representative immediately if you have any questions or concerns. An URGENT MEDICAL DEVICE RECALL notification letter dated 8/13/25 was sent to customers. ACTIONS REQUIRED 1. If leakage from the instrument is observed in the drip tray or surrounding the instrument, please follow your laboratory's leakage containment/safety protocol. If a protocol is running, wait for the run to finish, switch off the instrument and contact your local Leica Biosystems representative immediately. 2. If no leakage has been observed from the instrument, continue using it as normal. Conduct routine monitoring for leakage until the on-site inspection is completed. 3. Your local Leica Biosystems representative will contact you to schedule an onsite inspection of your instrument(s) for replacing the tubing and/or density meter manifold if needed. 4. Confirm Receipt of Recall Notice Please confirm receipt of this notice by signing and dating the attached Recall Acknowledgement Form by email to the following email ad

Distribution

US States: AZ, CA, FL, MN, NJ, NY, NC, OH, PA, TN, TX, UT, VA, and WI.

Quantity

77 units