QIAstat-Dx Respiratory SARS-CoV-2 Panel multiplexed nucleic acid real-time PCR test, Reference Number 691223

Recall: Z-1683-2023 · Initiated April 28, 2023

Recall

Recall Number
Z-1683-2023
Event Number
92298
Firm
Qiagen Sciences LLC
FEI Number
3003572099
Product Code
QLT
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
April 28, 2023
Posted
June 2, 2023
Address
19300 Germantown Rd, Germantown, MD, 20874-1415

Description

QIAstat-Dx Respiratory SARS-CoV-2 Panel multiplexed nucleic acid real-time PCR test, Reference Number 691223

Reason

Faulty cartridges in the lot could result in false test results.

Action

An Urgent Medical Device Correction notification letter dated April 2023 was sent to customers. Actions to be taken by customer/user Please check if you have remaining stock of LOT 175011350. Please DO NOT use it. Dispose of it immediately in accordance to your national and local safety and environmental regulations. Please contact QIAGEN Technical Services for a free-of-charge replacement. If you already used cartridges from this LOT, please identify the results obtained with cartridges from the affected SN. For results obtained with an affected SN, review the results as follows: o Review all results of the respective targets mentioned in this notice to exclude erroneous diagnosis and treatment, except in those cases where alternative confirmation was obtained. o Review any result for which tests were performed on clinical and/or epidemiological suspicion of the respective targets. Review this notice with your laboratory/medical director. IMPORTANT: Forward this information to all individuals and departments within your organization using the above listed kits. If you are not the end user, please forward this notice to the product end user. Complete the Acknowledgement of Receipt Form attached to this letter by 14 May 2023. Actions taken by QIAGEN QIAGEN was able to identify the root cause of the issue and has implemented immediate actions to ensure that the currently produced cartridges will not cause this issue moving forward. If you have any questions or concerns, please contact your local QIAGEN Technical Services Department through any of the following: Telephone: 800 362 7737 Email: [email protected]

Distribution

US Nationwide distribution in the states of CA, FL, IL, MD, MS, NY, OK, PA, TN.

Quantity

347 kits