FDA Recall Terminated

Accolade TMZF Plus Hip Stem; Hydroxylapatite coated, Titanium Plasma Spray. Hip stems are single use devices intended for cementless fixation. The products are intended for use in primary reconstruction of the proximal femur or revision of a previous total hip arthroplasty.

Recall: Z-1681-2009 · Initiated January 6, 2009

Recall

Recall Number
Z-1681-2009
Event Number
50901
Firm
Stryker Howmedica Osteonics Corp.
FEI Number
3003070421
Product Code
MEH
Status
Terminated
Root Cause
Process control
Initiated
January 6, 2009
Posted
July 22, 2009
Terminated
June 25, 2012
Address
325 Corporate Dr, Mahwah, NJ, 07430

Description

Accolade TMZF Plus Hip Stem; Hydroxylapatite coated, Titanium Plasma Spray. Hip stems are single use devices intended for cementless fixation. The products are intended for use in primary reconstruction of the proximal femur or revision of a previous total hip arthroplasty.

Reason

Product deviation : Specific lots of sprayed HA hip stems did not meet Stryker's Internal Material Specification for tensile bond strength and crystallinity.

Action

An "Urgent Product Recall" letter was sent via Federal Express on January 6, 2009 with return receipt to Stryker Branches/Agencies, Hospital Risk Management, Hospital Chief of Orthopaedics and Surgeons. The letter described the affected product, issue, potential hazards and risks. Stryker Branches and Hospital Risk Management were asked to examine their inventory for affected product. Stryker Representatives were instructed to contact customers to reconcile and return recalled product to Stryker Orthopaedics. All letters include a response form to be completed and returned indicating that they have been informed of the recall and the amount of product on hand, if any. Direct questions about the recall to Stryker Orthopaedics at 1-201-831-5718.

Distribution

Nationwide Distribution.

Quantity

1,628 total stems, all varieties.