FDA Recall Terminated

Wound Drain - Label reads in part "Elutia ***Hubless Coated Silicone Surgical Wound Drain***" Two different sizes, 10 units per box: 19 Fr Round Hubless part number 7110; 24 Fr Round Hubless part number 7118. AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system.

Recall: Z-1680-2014 · Initiated May 15, 2014

Recall

Recall Number
Z-1680-2014
Event Number
68305
Firm
Bacterin International, Inc.
FEI Number
3005031160
Product Code
OEI
Status
Terminated
Root Cause
Use error
Initiated
May 15, 2014
Posted
June 2, 2014
Terminated
June 2, 2014
Address
600 Cruiser Ln, Belgrade, MT, 59714-9719

Description

Wound Drain - Label reads in part "Elutia ***Hubless Coated Silicone Surgical Wound Drain***" Two different sizes, 10 units per box: 19 Fr Round Hubless part number 7110; 24 Fr Round Hubless part number 7118. AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system.

Reason

The recalling firm reported there was inadvertent omission of quarterly (Q4 - 2013 and Q1 - 2014) dose audits/verifications on the affected products. All of the affected products were appropriately sterilized

Action

Bacterin sent e-mail and FedEx letter with return receipt request to it's single consignee on May 15, 2014. Consignee is asked to separate the recalled product in a secure location for return to Bacterin International, Inc. Consignee is asked to complete tracking/verification form included in the letter/e-mail and call 1-888-886-9354 from 8:00AM to 5:00PM (Mountain Time) or send an e-mail to [email protected] to obtain a Return Authorization (RA) number.

Distribution

Distributed to one foreign consignee in Lebanon.

Quantity

390