FDA Recall Terminated

Acetaminophen Test System for use on the Roche/Hitachi, part number 03255379190, Roche Diagnostics Corporation, Indianapolis, IN. For the quantitative determinations of toxic levels of acetaminophen in human serum or plasma on Roche automated clinical analyzers.

Recall: Z-1679-2011 · Initiated December 7, 2010

Recall

Recall Number
Z-1679-2011
Event Number
57513
Firm
Roche Diagnostics Operations, Inc.
FEI Number
1823260
Product Code
LDP
Status
Terminated
Root Cause
Other
Initiated
December 7, 2010
Posted
March 16, 2011
Terminated
March 13, 2012
Address
9115 Hague Road, Indianapolis, IN, 46256-1025

Description

Acetaminophen Test System for use on the Roche/Hitachi, part number 03255379190, Roche Diagnostics Corporation, Indianapolis, IN. For the quantitative determinations of toxic levels of acetaminophen in human serum or plasma on Roche automated clinical analyzers.

Reason

The current product labeling for Acetaminophen does not provide specific interference information for bilirubin, hemoglobin or lipemia. Additional testing has been performed to quantify the impact of these interferences at different concentrations and different levels of acetaminophen. Low serum concentrations of acetaminophen were included in the study.

Action

The firm, Roche, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated December 07, 2010 to its customers. The letter described the product, problem, and action to be taken. The customers were instructed to be aware that endogenous interferents (i.e., bilirubin, hemolysis, and/or lipemia) may produce a falsely elevated value for samples containing concentrations less than 50 ug/ml of acetaminophen; complete and return the attached UMDC Faxback Form for Acetaminophen Assay via fax to 1-817-868-5343, and file this letter for future reference. The firm ask the customers to please consult with the physician or pathologist at their facility to determine specific clinical implications for their patients. The firm is also updating the product labeling to include information regarding the interference of icteric, hemolytic or lipemic samples. Please contact Roche Diagnostics Technical Support at 1-800-428-2336 if you have any questions about the information contained in this Urgent Medical Device Correction (UMDC).

Distribution

Nationwide distribution.

Quantity

8,668 units