Straight Inflow/Outflow Patient Line Kit. Model Number: 902-00037. Sterile, single-use, disposable set used with the Hyperthermia Pump. The Straight Inflow/Outflow Patient Line Kit (P/N 902-00037) consists of 4.4 Liter Reservoir (P/N 902-00034P), Heat Exchanger set (P/N 902-00006P) and Straight Inflow/Outflow Patient Line (P/N 902-00039P). The three components are packaged separately in individual pouches and boxes and placed in the same case for shipment. All three components are connected prior to use. The female quick connector that may be cracked is located on the Heat Exchanger set (P/N 902-00006P).
Recall
- Recall Number
- Z-1675-2025
- Event Number
- 96617
- Firm
- Belmont Instrument LLC
- FEI Number
- 1219702
- Product Code
- LGZ
- Status
- Open, Classified
- Root Cause
- Process design
- Initiated
- March 31, 2025
- Posted
- April 25, 2025
- Address
- 780 Boston Rd, Billerica, MA, 01821-5925
Description
Straight Inflow/Outflow Patient Line Kit. Model Number: 902-00037. Sterile, single-use, disposable set used with the Hyperthermia Pump. The Straight Inflow/Outflow Patient Line Kit (P/N 902-00037) consists of 4.4 Liter Reservoir (P/N 902-00034P), Heat Exchanger set (P/N 902-00006P) and Straight Inflow/Outflow Patient Line (P/N 902-00039P). The three components are packaged separately in individual pouches and boxes and placed in the same case for shipment. All three components are connected prior to use. The female quick connector that may be cracked is located on the Heat Exchanger set (P/N 902-00006P).
Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid infuser.
Belmont Medical notified consignees on about 03/31/2025 via "URGENT MEDICAL DEVICE CORECTION" FedEx or emailed letter. Consignees were instructed to locate affected product and to return the customer Response Form indicating if any affected units were on hand. Consignees were also instructed to follow instructions provided in the letter to inspect affected lots and if any units are identified as having cracks or leaks during priming, they should be discarded. Lot number of the set should be noted and reported to Belmont Medical Technologies. Belmont has indicated that they will replace any product found defective or at the request of the user any affected product that has not been used.
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, CT, DC, IL, IN, KS, KY, MA, ME, MN, NC, ND, NJ, NM, OH, OR, PA, SC, TN, TX, VA, VT, WA, WI, WV and the countries of Canada, Hungary, Korea (the Republic of), Latvia, Malaysia.
208 sets