FDA Recall
Terminated
AngioDynamics RITA Main Cable, 9 Pin to 14 Pin, REF 700-101892. RITA Main Cables are indicated for use with the RITA RF Generators and RITA electrosurgical devices.
Recall: Z-1675-2013
·
Initiated June 11, 2013
Recall
- Recall Number
- Z-1675-2013
- Event Number
- 65439
- Firm
- Angiodynamics, INC
- FEI Number
- 3017892510
- Product Code
- GEI
- Status
- Terminated
- Root Cause
- Packaging process control
- Initiated
- June 11, 2013
- Posted
- July 10, 2013
- Terminated
- June 12, 2015
- Address
- 1 Horizon Way, Manchester, GA, 31816-1749
Description
AngioDynamics RITA Main Cable, 9 Pin to 14 Pin, REF 700-101892. RITA Main Cables are indicated for use with the RITA RF Generators and RITA electrosurgical devices.
Reason
The RITA Main Cable product contains the incorrect Instructions for Use.
Action
Consignees were notified via certified mail on 06/11/2013 with an "Urgent Medical Device Field Safety Correction" notification. This included the correct Instructions for Use and the recommended actions.
Distribution
Worldwide Distribution, including the US, Austria, China, Italy, Russia and the UK.
Quantity
13 units