FDA Recall Terminated

Toshiba DUA-450F Kalare Radiographic Tilting Table.

Recall: Z-1675-2011 · Initiated July 15, 2010

Recall

Recall Number
Z-1675-2011
Event Number
56415
Firm
Toshiba American Medical Systems Inc
FEI Number
2020563
Product Code
IXR
Status
Terminated
Root Cause
Software design
Initiated
July 15, 2010
Posted
March 16, 2011
Terminated
April 20, 2012
Address
2441 Michelle Dr P.O. Box 2068, Tustin, CA, 92781-2068

Description

Toshiba DUA-450F Kalare Radiographic Tilting Table.

Reason

Top table lateral movement on their DUA-450F Kalare Table cannot be performed, or the table top lateral movement is performed at an extremely low speed. As a result, positioning of the patient cannot be performed by tabletop movement.

Action

Toshiba America Medical Systems (TAMS) issued Urgent Medical Device Correction notification dated July 8, 2010, to consignees via USPS return receipt mail. The letter contains a return reply form that is to be faxed to TAMS as proof of notification.

Distribution

US Nationwide

Quantity

238 systems