FDA Recall Terminated

Optiflex Lens Positioning Unit (LPU) Product Code: 10970, for use in vitreoretinal ophthalmic surgical procedures.

Recall: Z-1673-2009 · Initiated January 23, 2009

Recall

Recall Number
Z-1673-2009
Event Number
52461
Firm
Volk Optical Inc
FEI Number
1000122772
Product Code
KYB
Status
Terminated
Root Cause
Process design
Initiated
January 23, 2009
Posted
July 14, 2009
Terminated
April 12, 2012
Address
7893 Enterprise Dr, Mentor, OH, 44060-5309

Description

Optiflex Lens Positioning Unit (LPU) Product Code: 10970, for use in vitreoretinal ophthalmic surgical procedures.

Reason

Lens may fall onto eye surface during procedure. Ophthalmologic device had a loose set screw on the Ball-Seal bushing assembly. As a result, the set screw becomes disengaged from the set screw groove in the bushing assembly and the Lens Positioning Unit slides downward.

Action

Volk Optical sent a 'MEDICAL DEVICE RECALL' letter to their sole customer on 1/23/2009. The letter requests the return of the remaining two (2) units in the customer's possession (this firm experienced the device failure and reported the consumer complaint which prompted the recall; and had already returned the affected unit to the Volk for refund or replacement). The letter provides Volk's Fedex and UPS numbers for the customer to use to return the recalled units and also provides the E-mail address for the firm's contact person concerning this recall should additional assistance be required.

Distribution

USA distribution to a single customer located in MO.

Quantity

3 units