FDA Recall Terminated

Non-DEHP Microbore Admin Set with Non-Vented Spike and Needleless Injection Port Positive Pressure Package Assembly, product code: 340-4133, Method of Sterilization using Ethylene Oxide. The Curlin Ambulatory Infusion Pump System with disposable PVC Administration Tubing Sets and user accessories, provide a means for the volumetric delivery of fluids used in Parenteral, Enteral, Epidural, Subcutaneous and Intravenous applications.

Recall: Z-1671-2012 · Initiated May 18, 2012

Recall

Recall Number
Z-1671-2012
Event Number
61895
Firm
MOOG Medical Devices Group
FEI Number
1000117172
Product Code
FPK
Status
Terminated
Root Cause
Process design
Initiated
May 18, 2012
Posted
May 30, 2012
Terminated
February 22, 2013
Address
4314 Zevex Park Ln, Salt Lake City, UT, 84123-7881

Description

Non-DEHP Microbore Admin Set with Non-Vented Spike and Needleless Injection Port Positive Pressure Package Assembly, product code: 340-4133, Method of Sterilization using Ethylene Oxide. The Curlin Ambulatory Infusion Pump System with disposable PVC Administration Tubing Sets and user accessories, provide a means for the volumetric delivery of fluids used in Parenteral, Enteral, Epidural, Subcutaneous and Intravenous applications.

Reason

Moog Medical Device Group became aware on March 23, 2012 via customer complaint, of a reversed pump segment on an Administration Set. This malfunction could reverse the flow of fluid or medication backwards from what was intended.

Action

The firm, Moog Medical Device Group, sent an "Urgent Field Safety Notice" dated May 21, 2012 via certified letter and a press release direct to consignees/customers. The notice describes the product, problem and actions to be taken. The customers were instructed to do the following: If purchased by distributor -remove suspect product from inventory; contact your distributor to arrange replacement product and request a shipping return label as per the distributor's process and ship product to their location. If purchased by Moog-remove suspect product from inventory, contact Moog Customer Service at 1-800-970-2337, prompt #7 and Moog customer service will provide a call tag for convenient product return. For additional information or clarification, please contact the Post Market Surveillance Manager at (801) 264-1001, ext. 224 or via email at: [email protected].

Distribution

Nationwide distribution: USA including states of: AZ, CA, FL, IL, KS, MA, MI, MD, NC, ND, NY, OK, OH, and TX.

Quantity

3,880 sets.