FDA Recall Terminated

SIEMENS Uroskop Omnia Max ; a solid state detector fluoroscopic X-Ray system, primarily for urological applications (functional x-ray diagnostic, endourology and minimal invasive urology/surgery). The system,, which includes a radiologic/urologic treatment table, may be used for urological, gastroenterological and gynecological treatment, planning and diagnostic procedures including but not limited to: Querying and retrieving patient history information and/or previous diagnosis and images from other modalities, including X-ray examinations of the urogenital area, Ultrasound examinations, Endourological interventions, Percutaneous interventions, Laparoscopy, Application of fistula, Simple procedures, Extracorporeal shock wave lithotripsy, Uroflow/urodynamics, Pediatric radiological and therapeutic applications.

Recall: Z-1670-2015 · Initiated April 13, 2015

Recall

Recall Number
Z-1670-2015
Event Number
71115
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
JAA
Status
Terminated
Root Cause
Software design
Initiated
April 13, 2015
Posted
May 23, 2015
Terminated
March 20, 2017
Address
51 Valley Stream Pkwy, Malvern, PA, 19355-1406

Description

SIEMENS Uroskop Omnia Max ; a solid state detector fluoroscopic X-Ray system, primarily for urological applications (functional x-ray diagnostic, endourology and minimal invasive urology/surgery). The system,, which includes a radiologic/urologic treatment table, may be used for urological, gastroenterological and gynecological treatment, planning and diagnostic procedures including but not limited to: Querying and retrieving patient history information and/or previous diagnosis and images from other modalities, including X-ray examinations of the urogenital area, Ultrasound examinations, Endourological interventions, Percutaneous interventions, Laparoscopy, Application of fistula, Simple procedures, Extracorporeal shock wave lithotripsy, Uroflow/urodynamics, Pediatric radiological and therapeutic applications.

Reason

Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.

Action

A safety advisory notice, dated April 13, 2015, was sent to end users with affected devices to alert them about the issue and associated risks,and to provide instructions for avoiding potential risk. A new software version will be released to permanently resolve the issue.

Distribution

Nationwide Distribution.

Quantity

133 total