Synthes ProDisc-C 2.0mm Milling Bit, Synthes 2.0mm Milling Bit HXC, Synthes 2.0mm Milling Bit with 2.35mm x 33.5mm Shaft, and Synthes 2.0mm Milling Bit with Flat Coupling. Spinal surgical instrument: The Synthes ProDisc-C 2.0mm Milling Bits are used as a part of the milling option for keel preparation of the vertebral bodies. The milling bits are coupled to a power tool and interface with the milling guide, which is connected to a trial implant for the keel preparation step.

Recall: Z-1669-2014 · Initiated April 29, 2014

Recall

Recall Number
Z-1669-2014
Event Number
68190
Firm
Synthes, Inc.
FEI Number
3005180112
Product Code
GFG
Status
Terminated
Root Cause
Process control
Initiated
April 29, 2014
Posted
May 29, 2014
Terminated
May 19, 2016
Address
1302 Wrights Ln E, West Chester, PA, 19380-3417

Description

Synthes ProDisc-C 2.0mm Milling Bit, Synthes 2.0mm Milling Bit HXC, Synthes 2.0mm Milling Bit with 2.35mm x 33.5mm Shaft, and Synthes 2.0mm Milling Bit with Flat Coupling. Spinal surgical instrument: The Synthes ProDisc-C 2.0mm Milling Bits are used as a part of the milling option for keel preparation of the vertebral bodies. The milling bits are coupled to a power tool and interface with the milling guide, which is connected to a trial implant for the keel preparation step.

Reason

Certain lots of the ProDisc-C 2.0mm Milling Bits have an increased potential to break during use.

Action

DePuy Synthes sent a Recall letter, dated April 29, 2014, to affected customers. The letter identified the affected customers, problem and actions to be taken. The letter instructed customers to discontinue use and distribution of the device and to return affected devices to Synthes. For questions call 610-719-5450.

Distribution

US Nationwide Distribution

Quantity

6690