NOxBOXi Nitric Oxide Delivery System, NOXBOX-I, REQNOXBOX-I, Nitric Oxide Delivery Device

Recall: Z-1667-2023 · Initiated May 3, 2023

Recall

Recall Number
Z-1667-2023
Event Number
92171
FEI Number
3012796317
Product Code
MRN
Status
Terminated
Root Cause
Component change control
Initiated
May 3, 2023
Posted
June 9, 2023
Terminated
October 31, 2024
Address
NOXBOX LTD Unit 1080 & 1081 Galley Drive; Kent Science Park Sittingbourne United Kingdom

Description

NOxBOXi Nitric Oxide Delivery System, NOXBOX-I, REQNOXBOX-I, Nitric Oxide Delivery Device

Reason

Device may experience malfunctions due to misalignment of the check valve in the manifold. If the device fails, there is a risk of nitric oxide leak, oxygen gas leak, or earlier cylinder gas changeover than expected or not at all. In a worst case scenario, there may be an interruption in therapy which could lead to oxygen desaturation of the patient and associated complications.

Action

On May 3, 2023, the firm notified customers via URGENT MEDICAL DEVICE RECALL letters. Devices do not need to be removed from service unless they do not pass the start-up high-pressure leak test or it alarms during cylinder changeover and troubleshooting does not resolve the alarms. Customers should prioritize use of a device that is not affected, if available, when initiating therapy on a new patient. If there are no unaffected devices available, or if an affected device is already in use on a patient, follow troubleshooting measures as provided in the letter. Once new manifolds are available, NBL will correct or replace affected devices on a rolling basis. If you have any questions or require further assistance, please contact your sales representative or customer service at 833-NOX-VENT.

Distribution

Worldwide distribution - US Nationwide and the countries of Bahrain, Bolivia, Brazil, Ecuador, Finland, France, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Kuwait, Malaysia, Mexico, Netherlands, New Zealand, Panama, Saudi Arabia, Singapore, South Africa, Spain, Taiwan, Turkey, UAE, United Kingdom.

Quantity

548 (US); 208 (OUS)