FDA Recall Terminated

BD MAX Instrument, catalog #441916 Product Usage: The BD MAX System, used in conjunction with the appropriate extraction kits and reagents is capable of automated amplification and detection of target nucleic acid sequences for use in clinical, industrial, and research settings.

Recall: Z-1662-2013 · Initiated May 31, 2013

Recall

Recall Number
Z-1662-2013
Event Number
65431
Firm
Becton Dickinson & Co. BD Diagnostic Systems
FEI Number
1119779
Product Code
NJR
Status
Terminated
Root Cause
Component change control
Initiated
May 31, 2013
Posted
July 3, 2013
Terminated
August 19, 2013
Address
7 Loveton Circle, Sparks, MD, 21152-0999

Description

BD MAX Instrument, catalog #441916 Product Usage: The BD MAX System, used in conjunction with the appropriate extraction kits and reagents is capable of automated amplification and detection of target nucleic acid sequences for use in clinical, industrial, and research settings.

Reason

Electrical power module installed on some in vitro diagnostic instruments does not meet industry standards as labeled.

Action

Becton Dickinson (BD) notified their customers in the US by phone on May 31, 2013 and by an Urgent Field Correction Action letter dated May 2013. Customers were advised of the affected product, problem and actions to be taken. The letter informed customers that an onsite visit by a BD Service representative is planned to perform a system check so to determine if the instrument is affected. BD's field service team plans to immediately replace the module on site if the affected part is found. For questions contact BD Technical Services Department at 1-800-638-8663.

Distribution

Worldwide Distribution - USA Nationwide and in the countries of: Japan, Singapore, Hong Kong, China, Australia, Europe, and Chile.

Quantity

71 instruments