FDA Recall Open, Classified

Venous Hardshell Cardiotomy Reservoir, used during cardiopulmonary bypass Model Item No. VKMO 10000-USA 701070441 VKMO 11000-USA 701070445

Recall: Z-1661-2023 · Initiated May 18, 2023

Recall

Recall Number
Z-1661-2023
Event Number
92006
Firm
Maquet Medical Systems USA
FEI Number
3008355164
Product Code
DTZ
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
May 18, 2023
Posted
May 30, 2023
Address
45 Barbour Pond Dr, Wayne, NJ, 07470-2094

Description

Venous Hardshell Cardiotomy Reservoir, used during cardiopulmonary bypass Model Item No. VKMO 10000-USA 701070441 VKMO 11000-USA 701070445

Reason

Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.

Action

On May 18, 2023, the firm notified customers of the recall via an Urgent Medical Device Removal letter. Actions to be taken by the customer: If an affected device is already in use, please use according to normal practices. Monitor the patient for signs and symptoms of inflammation, infection, sepsis, and ischemia. If detected, treat according to clinical protocols. Please remove any affected product in inventory from areas of use. Unopened and unexpired affected product is eligible for credit upon return. Product Return: Please contact Getinge Customer Service at 1-888-943-8872 option 2 between the hours of 6:00 AM and 5:00 PM Pacific Standard Time to request a return authorization (RMA) and shipping instructions to return any affected product. Distributors who have shipped affected products to customers should forward the notification to customers' attention for appropriate action. Maquet Cardiopulmonary GmbH is working with all possible urgency on the redesign and revalidation of a packaging solution, which is anticipated to be available by Q4 2023.

Distribution

Nationwide domestic distribution

Quantity

0 (US)