FDA Recall Open, Classified

Philips Brilliance CT Big Bore, Software Version 4.8.0.10430. Computed Tomography X-Ray System.

Recall: Z-1660-2025 · Initiated April 1, 2025

Recall

Recall Number
Z-1660-2025
Event Number
96667
Firm
Philips North America Llc
FEI Number
3016618143
Product Code
JAK
Status
Open, Classified
Root Cause
Software design
Initiated
April 1, 2025
Posted
April 23, 2025
Address
222 Jacobs St, Cambridge, MA, 02141-2289

Description

Philips Brilliance CT Big Bore, Software Version 4.8.0.10430. Computed Tomography X-Ray System.

Reason

Unintended motion issues related to Interventional Control that may lead to collision of the Gantry/table with the operator or patient.

Action

An "URGENT Medical Device Correction" Notice dated 4/1/25 was mailed to consignees. The notice instructs consignees on how to identify if their unit is affected by this recall. Consignees are instructed to continue use of affected devices by following the recommended actions outlined in the notice and Appendix A. Consignees are asked to return the completed URGENT Medical Device Correction Response Form within 30 days of notification of this recall event via email: [email protected]. The recall notification is to be circulated among users and retained near affected devices until the update is installed. Philips will be contacting consignees to schedule a Field Service Engineer to visit affected sites and install the update (FCO72800825). Consignees with any questions are to call the Customer Care Solutions Center at 1-800-722-9377.

Distribution

Worldwide distribution: US (Nationwide); and OUS (Foreign) to countries of: Algeria, Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, China, Czech Republic, Estonia, France, Germany, Honduras, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Lebanon, Libya, Luxembourg, Mauritius, Mexico, Monaco, Morocco, Myanmar, Netherlands, New Zealand, Norway, Palestine, Panama, Philippines, Poland, Portugal, Puerto Rico, R¿union, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, T¿rkiye, Turkmenistan, United Kingdom, UAE & Yemen.

Quantity

630 units