FDA Recall Terminated

***REF SC2101***eMax 2 Plus Console, Single Port, with Irrigation***SN G17307901802*** Product Usage: The RPM of the Motor Handpiece may be less than 80,000 RPM (approximately 72,000 RPM) in the free running state when using the optional Hand Control. The low Motor Handpiece RPM would only be experienced in the Free running state (Hand Control fully depressed and no load on the cutting bur). This condition would not affect the power output of the Handpiece in the loaded working range (typically 50,000 to 70,000 RPM.

Recall: Z-1658-2013 · Initiated June 24, 2011

Recall

Recall Number
Z-1658-2013
Event Number
64973
Firm
The Anspach Effort, Inc.
FEI Number
1045834
Product Code
NAV
Status
Terminated
Root Cause
Process control
Initiated
June 24, 2011
Posted
July 2, 2013
Terminated
December 24, 2013
Address
4500 Riverside Drive, Palm Beach Gardens, FL, 33410-4235

Description

***REF SC2101***eMax 2 Plus Console, Single Port, with Irrigation***SN G17307901802*** Product Usage: The RPM of the Motor Handpiece may be less than 80,000 RPM (approximately 72,000 RPM) in the free running state when using the optional Hand Control. The low Motor Handpiece RPM would only be experienced in the Free running state (Hand Control fully depressed and no load on the cutting bur). This condition would not affect the power output of the Handpiece in the loaded working range (typically 50,000 to 70,000 RPM.

Reason

The RPM of the Motor Handpiece may be less than 80,000 RPM.

Action

Anspach Effort sent an "URGENT: Medical Device Correction" letter dated June 24, 2011 to their customers via mail. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete a reply form promptly confirming receipt of their letter and return it by fax to the number provided on the form. If any of the affected products were distributed to other services or facilities, please forward this information as appropriate. For questions contact Anspach Customer Support at 800-327-6887 or email [email protected].

Distribution

Worldwide Distribution - USA Nationwide including the states of: AZ, CA, FL, KY, ME, MI, MO, OH, PA, TN, TX , VA, and the countries of: Australia, Egypt, Italy, Japan, Switzerland, Taiwan, and Turkey;

Quantity

61 units