Unicel DxH 800 Coulter Cellular Analysis System. Beckman Coulter, Inc. Brea, CA. Quantitative, automated hematology analyzer and for in vitro diagnostic use in screening patient populations found in clinical laboratories.
Recall
- Recall Number
- Z-1658-2010
- Event Number
- 55176
- Firm
- Beckman Coulter Inc.
- FEI Number
- 2050012
- Product Code
- GKZ
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- February 4, 2010
- Posted
- May 26, 2010
- Terminated
- September 2, 2014
- Address
- 200 S Kraemer Blvd, Brea, CA, 92821-6208
Description
Unicel DxH 800 Coulter Cellular Analysis System. Beckman Coulter, Inc. Brea, CA. Quantitative, automated hematology analyzer and for in vitro diagnostic use in screening patient populations found in clinical laboratories.
Issue 1: The system will substitute or omit the characters # @ 1 \ ] ' ( I ) - when the instrument is configured in a language other than English, Chinese or Japanese. Example: Using the character @, a substitution in the string '123@456' would result in '123#456' and an omission would result in '123456'. Issue2: The characters 7 " and * (question mark, quotation mark, and asterisk) used as part
Beckman Coulter, Inc. issued "Urgent: Product Corrective Action" notifications dated February 12, 2010. Consignees were informed of the affected product and instructed on proper actions. The firm also asked that the enclosed response forms be returned. For further information, contact your local Beckman Coulter Representative or call 1-800-526-7694.
Distributed Nationwide and Canada.
56 units