Welch Allyn AED 10 automatic external defibrillator and MRL Jumpstart (collectively "AED 10")
Recall
- Recall Number
- Z-1656-2009
- Event Number
- 51923
- Firm
- Welch Allyn Protocol, Inc
- FEI Number
- 3023750
- Product Code
- MKJ
- Status
- Terminated
- Root Cause
- Labeling design
- Initiated
- April 24, 2009
- Posted
- July 6, 2009
- Terminated
- August 11, 2011
- Address
- 8500 Sw Creekside Pl, Beaverton, OR, 97008
Description
Welch Allyn AED 10 automatic external defibrillator and MRL Jumpstart (collectively "AED 10")
Possible for users to misunderstand signals. The original product labeling told users that a flashing low battery indicator signals that the battery is low and requires changing soon. If a user experiences a flashing low battery status indicator and does not understand that the device can continue to be used, the user might choose not to continue operating the device, thereby causing therapy to
The firm has developed expanded labeling for users and will revise the user manuals for software versins 2.03, 2.04, 2.05, and 2.06. Customers receiving the updated manuals will be sent a cover letter directing them to discard the original user manual and replace it with the updated manual. The April 24, 2009, cover letter will highlight the clarification of directions for use regarding what to do when a low battery indicator is displayed.
Worldwide distribution: USA, Guam, Puerto Rico, Canada, Mexico, Andorra, UAE, Afghanistan, Australia, Azerbaijan, Belgium, Belarus, Bermuda, Bahrain, Brazil, British West Indies, Switzerland, China, Cyprus, Czech Republic, Germany, Denmark, Algeria, Egypt, Spain, France, French Guiana, French Lesser Antilles, Greece, Haiti, Hong Kong, Hungary, Ireland, Israel, India, Iraq, Italy, Japan, Kenya, Kyrgyz, Lebanon, Lithuania, Luxembourg, Latvia, Monaco, Malaysia, Netherlands, New Zealand, Phillipines, Pakistan, Poland, Portugal, Reunion Island, Russia, Saudi Arabia, Singapore, South Korea, Taiwan, United Kingdom, Uraguay, Yemen, and South America.
46740 units