FDA Recall Terminated

Brilliance CT 6 Air. Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

Recall: Z-1655-2015 · Initiated March 16, 2015

Recall

Recall Number
Z-1655-2015
Event Number
71177
Firm
Philips Medical Systems (Cleveland) Inc
FEI Number
1525965
Product Code
JAK
Status
Terminated
Root Cause
Employee error
Initiated
March 16, 2015
Posted
May 22, 2015
Terminated
February 22, 2017
Address
595 Miner Rd, Cleveland, OH, 44143-2131

Description

Brilliance CT 6 Air. Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

Reason

The firm was informed that while raising the patient couch on the system to perform an exam, the couch unexpectedly descended to the lowest point without being commanded to do so.

Action

On 3-18-2015 the firm sent Urgent Field Safety Notices to their customers.

Distribution

Distributed in China.

Quantity

153 units total