Philips IntelliVue Clinical Information Portfolio Software Version D.0, for patient data collection, storage, and management.

Recall: Z-1654-2009 · Initiated May 5, 2009

Recall

Recall Number
Z-1654-2009
Event Number
51979
Firm
Philips Healthcare Inc.
FEI Number
1218950
Product Code
DXJ
Status
Terminated
Root Cause
Software design
Initiated
May 5, 2009
Posted
July 7, 2009
Terminated
February 6, 2017
Address
3000 Minuteman Road, Andover, MA, 01810

Description

Philips IntelliVue Clinical Information Portfolio Software Version D.0, for patient data collection, storage, and management.

Reason

Error in therapy scheduling: If scheduled orders are documented and edited prior to the orders' original schedule time, ICIP can create a second instance of the same order marked as pending. A second instance of an order may result in a patient receiving duplicate therapy or patients may not receive their scheduled therapies.

Action

Philips notified customers by Urgent Medical Device Correction Notification letter on May 6, 2009. The Customer notifications in the United States will be delivered via UPS. Customer notifications outside the US will be managed by Philips representatives in each affected geography. Customers will be asked to follow the guidelines in the Action to be taken by Customer/User section of the Urgent Medical Device Correction Notice until they receive their software corrections. The corrections will be in the form of software upgrades and will be performed by Philips service personnel. On May 27, 2009, Philips issued the second Urgent Medical Device Correction notification. While customers await the correction of their devices, customers are given the following instructions: Clinicians should not rely solely on the worklist within lCIP to determine actionable orders for the patient until their software is upgraded. Clinical users should always use the medication administration record (MAR) and the administration record in determining the care their patients receive in addition to the critical care worklist.

Distribution

Worldwide distribution, including USA, Australia, Belgium, China, Denmark, France, Germany, Ireland, Italy, Kenya, Malaysia, Netherlands, Singapore, Spain, Switzerland, Taiwan, UAE, and UK.

Quantity

105 units