FDA Recall Open, Classified

ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.

Recall: Z-1653-2026 · Initiated March 10, 2026

Recall

Recall Number
Z-1653-2026
Event Number
98675
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
OWB
Status
Open, Classified
Root Cause
Radiation Control for Health and Safety Act
Initiated
March 10, 2026
Posted
April 7, 2026
Address
40 Liberty Blvd, Malvern, PA, 19355-1418

Description

ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.

Reason

During patient examinations, the system may sporadically display a major increase of the x-ray dose applied to the patient. Despite the reported dose showing significantly higher, the actual applied x-ray dose levels are in accordance with the system settings and imaging conditions, as well as the applicable regulatory requirements.

Action

Siemens Medical Solutions USA, Inc. has initiated a Customer Advisory Notice, dated March 9, 2026, to all affected customers.

Distribution

Worldwide - US Nationwide distribution.

Quantity

U.S. 758, OUS 2077