FDA Recall Terminated

ADVIA Centaur Systems TSH3 Ultra Ready Pack, 100 test Catalog Number: 06491072; Siemens Material Number: 10282378. Intended for the invitro diagnostic use in quantitative determination of various assays on the ADVIA Centaur and ADVIA Centaur XP analyzers.

Recall: Z-1653-2014 · Initiated April 10, 2014

Recall

Recall Number
Z-1653-2014
Event Number
67953
Firm
Siemens Healthcare Diagnostics, Inc
FEI Number
1219913
Product Code
JLW
Status
Terminated
Root Cause
Device Design
Initiated
April 10, 2014
Posted
May 23, 2014
Terminated
May 27, 2016
Address
333 Coney St, East Walpole, MA, 02032-1516

Description

ADVIA Centaur Systems TSH3 Ultra Ready Pack, 100 test Catalog Number: 06491072; Siemens Material Number: 10282378. Intended for the invitro diagnostic use in quantitative determination of various assays on the ADVIA Centaur and ADVIA Centaur XP analyzers.

Reason

Samples containing fluorescein may show interference with the ADVIA Centaur Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays.

Action

Siemens Healthcare Urgent Medical Device Correction (UMDC) was mailed to all affected Siemens Healthcare Diagnostics customers in the United States on April 11, 2014 for customer receipt on April 14, 2014. This notice informed the customers of fluorescein interference with the ADVIA Centaur Systems TSH3-Ultra and Vitamin D Total assays. This notice instructs customers that the TSH3-Ultra and Vitamin D assays can continue to be used for testing any sample that does not contain fluorescein. However, testing blood drawn from a patient, without allowing clearance of fluorescein is not recommended. This notice should be reviewed with the facility's laboratory or medical director. B. An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on April 10, 2014. This notice informed the customer of fluorescein interference with the ADVIA Centaur Systems TSH3-Ultra , Vitamin D Total and ADVIA Centaur BRAHMS Procalcitonin assays. This notice instructs customers that the TSH3-Ultra, Vitamin D and Procalcitonin assays can continue to be used for testing any sample that does not contain fluorescein. However, testing blood drawn from a patient, without allowing clearance of fluorescein is not recommended. This notice should be reviewed with the facilitys laboratory or medical director. (Note: The ADVIA Centaur BRAHMS Procalcitonin assay is not sold in the US.)

Distribution

Worldwide Distribution-USA (nationwide) and the countries of Afghanistan Algeria Armenia Austria Bahrain Belgium Bulgaria Burkina Faso Canary Islands Croatia Cyprus Czech Republic Denmark Estonia Finland France Germany Great Britain Greece Hungary Iceland Iran Iraq Ireland Israel Italy Jordan Kazakhstan Latvia Lebanon Lithuania Luxembourg Macedonia Malta Netherlands Norway Oman Pakistan Poland Portugal Qatar Romania Russian Fed. Saudi Arabia Serbia Slovakia Slovenia Spain Sweden Switzerland Turkey Unit.Arab Emir. Vatican Yemen.

Quantity

22,770 kits