FDA Recall Terminated

ABL700 Series analyzer - Human Blood Analyzer (Laboratory Device).

Recall: Z-1653-2008 · Initiated January 18, 2008

Recall

Recall Number
Z-1653-2008
Event Number
46686
Firm
Radiometer America Inc
FEI Number
1523456
Product Code
CHL
Status
Terminated
Root Cause
Software design
Initiated
January 18, 2008
Terminated
February 24, 2012
Address
810 Sharon Dr, Westlake, OH, 44145-1598

Description

ABL700 Series analyzer - Human Blood Analyzer (Laboratory Device).

Reason

Reported calibration errors are not displayed on the analyzers parameter bar screen or on the patient results when the calibration number is between 32768 - 65535 and 98303 - 131070. This may cause the operator to perform measurements and obtain questionable results.

Action

On February 28, 2008, a Field Action Notes Letter was issued advising them about a Mandatory Software Update. Customers were notified of the problem by fax and letter (Fed Ex). The Recall Letter instructed the user to disable the parameter bar feature and only rely on the traffic light system indicator which properly indicates system errors. The letter includes instructions to temporarily fix the problem and notice of a update of the software module that updates the parameter bar. If you have any questions contact Vincent Sigmund at (440)-871-8900 ext. 209.

Distribution

Worldwide Distribution - USA and Canada - Products were sold in all fifty (50) US States. including twenty-nine 29 government accounts and 138 Canadian accounts.

Quantity

2481 Units