Hitachi ECLOS Computed Tomography (CT) Scanners, Software Version: 2.11 or less.

Recall: Z-1652-2009 · Initiated April 14, 2009

Recall

Recall Number
Z-1652-2009
Event Number
52430
Firm
Hitachi Medical Systems America Inc
FEI Number
1528028
Product Code
JAK
Status
Terminated
Root Cause
Software design
Initiated
April 14, 2009
Posted
July 7, 2009
Terminated
September 21, 2012
Address
1959 Summit Commerce Park, Twinsburg, OH, 44087

Description

Hitachi ECLOS Computed Tomography (CT) Scanners, Software Version: 2.11 or less.

Reason

Incorrect scale on image: A software error can occur if two (2) different Field-of-View (FOV) settings are registered in a protocol using Multi-Reconstruction. This could cause the image generated to be displayed with the wrong scale factor rendering anatomical measurements incorrect.

Action

HMSA issued an URGENT - DEVICE CORRECTION notification to their customers on 4/14/2009. The notification describes the potential software error and the condition under which these image distortions can occur. The notification goes on to provide the customer with a work-around procedure which they can use to avoid encountering this issue with their current software configuration(s). The letter states that the firm will be installing updated software on all affected systems beginning in June, 2009, and that a Hitachi Service Rep. will be contacting each customer and making arrangements to come out and install a software upgrade on the affected system(s) at no cost to the customer. Lastly, the letter provides a telephone # to the firm's Applications Helpline: 1-877-228-2777, which the customer can use to obtain additional information, if desired.

Distribution

Nationwide distribution, including Puerto Rico (PR).

Quantity

4 units