FDA Recall Open, Classified

Puracol and Puracol Plus Collagen Wound Dressings: MSC8522 DRESSING,COLLAGEN,PURACOL,STRL,2" X 2" MSC8622EP DRESSING,COLLAGEN,PURACOL PLUS, 2X2.2"

Recall: Z-1650-2026 · Initiated February 10, 2026

Recall

Recall Number
Z-1650-2026
Event Number
98484
FEI Number
1417592
Product Code
KGN
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
February 10, 2026
Posted
March 27, 2026
Address
3 Lakes Dr, Medline Industries, LP, Northfield, IL, 60093-2753

Description

Puracol and Puracol Plus Collagen Wound Dressings: MSC8522 DRESSING,COLLAGEN,PURACOL,STRL,2" X 2" MSC8622EP DRESSING,COLLAGEN,PURACOL PLUS, 2X2.2"

Reason

Elevated endotoxin levels identified in affected lots. Issue may elicit strong inflammatory response if used, including local tissue irritation, delayed wound healing, or systemic effects such as fever and hypotension. In severe cases, septic shock may result.

Action

Firm began notifying customers on 2/10/26 via "Urgent Medical Device Recall" letters. Customers were instructed to immediately locate and quarantine all affected product. Affected product must be destroyed. Customer should respond to the firm to confirm quantity of affected product in inventory. The customer will receive credit from the firm once the response is submitted. Distributors should notify any of their affected downstream accounts. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].

Distribution

Domestic distribution nationwide US. International distribution to Mexico and Panama.

Quantity

106,621 US; 2,350 OUS