FDA Recall Terminated

NeuViz 64 Multi-Slice CT Scanner System (consist if two variants: NeuViz 64e, NeuViz 64i)

Recall: Z-1650-2016 · Initiated May 2, 2015

Recall

Recall Number
Z-1650-2016
Event Number
73832
Firm
Neusoft Medical Systems Co., Ltd. NO.
FEI Number
3003575350
Product Code
JAK
Status
Terminated
Root Cause
Device Design
Initiated
May 2, 2015
Terminated
November 4, 2016
Address
16 Shiji Road, Hunnan Industrial Area Shenyang China

Description

NeuViz 64 Multi-Slice CT Scanner System (consist if two variants: NeuViz 64e, NeuViz 64i)

Reason

It was found by R&D in April, 2014 that there was a defect in NeuViz 64 system software version 1.0.5+P09. The icons of side decubitus didn't meet the patient position description, when the scanner position was set to "Right" in the system setting - scanner options.

Action

An upgraded software version (1.0.5.2619+P11) was released on May 7, 2015. The FCO was released to the service engineers of Neusoft Medical Systems USA, which is subsidiary of the reporting firm, and the engineers performed FCO for each affected device together with the distributors' engineers. The corrections (FCOs to upgrade software) on all the affected systems in the USA have been completed. At present (4/27/16) there is no longer this issue on the systems, so no further actions are needed for customers/distributors. However, the recalling firm shall inform the direct accounts of this fix before May 5, 2016. The recalling firm will submit the communication evidence ASAP once they have evidence of the communication. For further questions, please call (281) 453-1205.

Distribution

US Distribution - Including Puerto Rico and the states of IL, CT, SC, NE

Quantity

7 units