FDA Recall Terminated

Covidien Kangaroo Connect Enteral Feeding Pump Item Number: 584400 (OUS) Intended to provide enteral nutrition to a patient for hospital and acute care settings.

Recall: Z-1649-2016 · Initiated April 1, 2016

Recall

Recall Number
Z-1649-2016
Event Number
73669
Firm
Medtronic
FEI Number
1219930
Product Code
LZH
Status
Terminated
Root Cause
Device Design
Initiated
April 1, 2016
Terminated
May 18, 2018
Address
60 Middletown Ave, North Haven, CT, 06473-3908

Description

Covidien Kangaroo Connect Enteral Feeding Pump Item Number: 584400 (OUS) Intended to provide enteral nutrition to a patient for hospital and acute care settings.

Reason

Kangaroo Connect Feeding Pump Occlusion alarms fail to alarm

Action

Medtronic(formerly) Covidien notified accounts via Urgent Medical Device Recall letter, dated 4/1/16, that specific serial numbers with software versions 1.12 and 1.13 have a software anomaly. The firm shipped replacement units with software V 1.14. Affected devices must be returned to the firm once the replacement units have been received. A Recalled Product Return Form required to be completed and returned. Questions or concerns, contact Medtronic representative or Customer Service at (800) 882-5878.

Distribution

Worldwide Distribution -- USA, Australia, Canada, France, and Singapore.

Quantity

145 units