FDA Recall Terminated

Bio Rad brand Autoimmune EIA Anti-dsDNA Test Kit, 576 Tests, Catalog No. 576DS, Distributed by and Manufactured by: Bio-Rad Laboratories, Inc. Hercules, CA The Bio-Rad Autoimmune EIA Anti-dsDNA Test is a qualitative enzyme immunoassay (EIA) intended to screen for the presence of dsDNA antibodies in human serum as an aid in the diagnosis of systemic lupus erythematosus (SLE). For In Vitro Diagnostic Use.

Recall: Z-1648-2011 · Initiated March 2, 2010

Recall

Recall Number
Z-1648-2011
Event Number
56044
Firm
Bio-Rad Laboratories Inc
FEI Number
2915274
Product Code
LRM
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
March 2, 2010
Posted
March 14, 2011
Terminated
March 15, 2011
Address
4000 Alfred Nobel Dr, Hercules, CA, 94547-1803

Description

Bio Rad brand Autoimmune EIA Anti-dsDNA Test Kit, 576 Tests, Catalog No. 576DS, Distributed by and Manufactured by: Bio-Rad Laboratories, Inc. Hercules, CA The Bio-Rad Autoimmune EIA Anti-dsDNA Test is a qualitative enzyme immunoassay (EIA) intended to screen for the presence of dsDNA antibodies in human serum as an aid in the diagnosis of systemic lupus erythematosus (SLE). For In Vitro Diagnostic Use.

Reason

The firm received customer complaints of increased positivity rates associated with the Anti-dsDNA test kits.

Action

The firm, Bio-Rad, sent an "URGENT: MEDICAL DEVICE RECALL NOTIFICATION" letter dated March 2, 2010 to all customers. The letter described the product, problem and action to be taken by the customers. The customers were instructed to examine their inventory and quarantine the affected kit lots; discard all affected product in accordance with all local, state and federal disposal procedures; complete and return the enclosed MEDICAL DEVICE RECALL RESPONSE FORM via fax to: (510) 741-3954, attn: Bio-Rad CSD Regulatory Affairs Department, and to please re-evaluate all positive patient samples and retest if necessary. If you have any questions, contact Bio-Rad CSD Regulatory Affairs Department at 510-724-7000.

Distribution

Worldwide distribution: USA and countries including: Canada, Carribean, Chile, France, Germany, Hong Kong, India, Italy, Japan, New Zealand, Panama, Singapore, South Korea and United Kingdom.