FDA Recall Terminated

Siemens Medical Solutions USA, Inc., Symbia S Series SPECT System; Single-Photon Emission Computed Tomography; Part number 8717741.

Recall: Z-1648-2009 · Initiated March 11, 2009

Recall

Recall Number
Z-1648-2009
Event Number
51483
Firm
Siemens Medical Solutions USA, Inc.
FEI Number
1000514697
Product Code
KPS
Status
Terminated
Root Cause
Software design
Initiated
March 11, 2009
Posted
July 7, 2009
Terminated
September 11, 2010
Address
2501 Barrington Rd, Hoffman Estates, IL, 60195-2061

Description

Siemens Medical Solutions USA, Inc., Symbia S Series SPECT System; Single-Photon Emission Computed Tomography; Part number 8717741.

Reason

Patient injury can occur when the patient is lying on the patient bed and the automatic collimator exchanger is undergoing a collimator change operation at the launch of an acquisition workflow. Detector motions required for this change can cause collision with the patient or pallet if it is in brain scan position, i.e. the patient or pallet is into the field of view.

Action

Siemens sent Urgent Field Correction Recall letters dated 1/20/09 to all users of the Symbia S and Symbia T Systems. The accounts were informed of the potential injury to patients while lying on the patient bed and the automatic collimator exchanger is undergoing a collimator change operation at the launch of an acquisition workflow. The customers were advised to perform all necessary collimator changes before the patient is positioned on the patient bed, and to examine each of their workflows that include a Tomo Acquisition activity and make sure the Auto Collimator Change check box on the Camera Parameters Tab is not checked. They also were advised that a Siemens Customer Service Representative would contact them within 60 days to schedule a corrective software upgrade free of charge. Any questions were directed to contact Siemens Medical Solutions USA, Inc. at 1-800-767-2313 (USA) for assistance. The accounts were requested to complete the Mandatory Fax-Back Form acknowledging receipt and understanding of the letter.

Distribution

Worldwide distribution, including USA, Australia, Austria, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Ireland, Italy, Japan, Korea, Kuwait, Netherlands, Norway, Poland, Portugal, Qatar, Saudi Arabia, South Africa, Spain, Sweden and Switzerland.

Quantity

152 units