FDA Recall Open, Classified

BD Alaris" System with Guardrails" Suite MX with Point of Care Unit

Recall: Z-1645-2026 · Initiated February 11, 2026

Recall

Recall Number
Z-1645-2026
Event Number
98250
Firm
CareFusion 303, Inc.
FEI Number
2016493
Product Code
FRN
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
February 11, 2026
Posted
March 26, 2026
Address
10020 Pacific Mesa Blvd, San Diego, CA, 92121-4386

Description

BD Alaris" System with Guardrails" Suite MX with Point of Care Unit

Reason

Due to product labeling not indicating which Transport Layer Security (TLS) version that must be enabled on hospital network to ensure secure Wi-Fi communication with Point-of-Care Units

Action

On 02/11/2026, the firm sent via US Postal mail, an "URGENT MEDICAL DEVICE PRODUCT CORRECTION" letter to customers informing them that labeling does not indicate which Transport Layer Security (TLS) versions must be enabled on the hospital network to ensure secure Wi-Fi communication with the Alaris and BD Alaris Point-of-Care Unit (PCU) when using enterprise-level Wi-Fi security. Without clear labeling, hospitals may unknowingly disable TLS 1.0, resulting in network connectivity failure and memory leaks which could lead to potential System Error. Customers/clinicians are instructed to: 1. Ensure the PCU is connected to the hospital network by viewing the wireless indicator icon on the front of the PCU. If the icon is illuminated or blinking, the PCU is connected. If it is not illuminated, the PCU is not connected. 2. If icon is not illuminated, follow established hospital protocols for wireless network downtime. 3. If a network connectivity failure persists, System Error may occur on the PCU. This error is indicated by a high priority alarm and red blinking arrow above the SYSTEM ON button. The customer must power down the PCU, then power on again. Any infusion that was ongoing at the time of System Error must be reprogrammed. 4. If using enterprise-level network security, customers/IT Security should be aware of which versions of TLS must be enabled to ensure connectivity. Refer to Appendix A for additional network administration guidance and compatibility. 5. Disseminate and post this notice within their facility to ensure that all concerned personnel, including IT and network connectivity owners or administrators, are notified and aware of this issue. For Questions/Assistance - contact the Recall Support Center at: Phone: 1-888-562-6018 Phone Hours: 7:00am to 4:00pm PT, Monday- Friday Email: [email protected]

Distribution

Worldwide distribution: U.S.(Nationwide) to states of: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP. MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY O.U.S. (foreign) to countries of: Australia, Belgium, Bahrain, Botswana, Canada, China, Cuba, Germany, India, Israel, Italy, Japan, Korea, Kuwait, Malaysia, Mexico, Netherlands, New Zealand, Pakistan, Philippines, Qatar, Saudi Arabia, Singapore, South Africa, Switzerland, Taiwan, Turkey, United Arab Emirates, and United Kingdom.

Quantity

1,025,567