FDA Recall Terminated

XiO Radiation Treatment Planning System, Versions 4.30.00-4.64.00. Used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed. The system will calculate and display, both on-screen and in hard-copy, either two- or three-dimensional radiation dose distributions within a patient for a given treatment plan set-up.

Recall: Z-1644-2012 · Initiated April 30, 2012

Recall

Recall Number
Z-1644-2012
Event Number
61746
Firm
Computerized Medical Systems Inc
FEI Number
1937649
Product Code
MUJ
Status
Terminated
Root Cause
Software design
Initiated
April 30, 2012
Posted
May 22, 2012
Terminated
April 1, 2014
Address
13723 Riverport Dr, Ste 100, Maryland Heights, MO, 63043-4819

Description

XiO Radiation Treatment Planning System, Versions 4.30.00-4.64.00. Used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed. The system will calculate and display, both on-screen and in hard-copy, either two- or three-dimensional radiation dose distributions within a patient for a given treatment plan set-up.

Reason

XiO will underestimate the dose delivered when the closed leaf gap is set to 0mm on XiO and a value other than 0mm is set on the accelerator.

Action

Computerized Medical Systems, Inc. sent an Important Safety Notice the week of April 30, 2012, by regional support managers via e-mail, fax, or regular mail to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. North American customers were instructed to go to www.elekta.com, select the SupportPlus Login at the top of the page, enter their portal, select Downloads/Updates to download the latest patches. For Customers outside North America, go to www.elekta.com, select the SOFTWARE button on the left, Treatment Planning Software, then Contact TPS Support tab, Download TPS Software Patches. ............................................................. Customers were also instructed to distribute this notice to any and all users of CMS software at their organization who are potentially affected by this issue. For questions regarding this recall call 408-380-8023.

Distribution

Worldwide Distribution - USA (nationwide) and the country of Canada

Quantity

1,652 sites received the versions