Medtronic Prestige LP(TM) Cervical Disc System with Streamlined Instruments, DRILL GUIDES: a) REF 6975205, 5mm b) REF 6975206, 6mm c) REF 6975207, 7mm Product Usage: The PRESTIGE LP reusable instruments are intended for the surgical implantation of the PRESTIGE LP Cervical Disc. Medtronic reusable instruments must be sterilized prior to initial use and must be cleaned and sterilized prior to each re-use.
Recall
- Recall Number
- Z-1641-2018
- Event Number
- 79667
- Firm
- Medtronic Sofamor Danek USA Inc
- FEI Number
- 1000477302
- Product Code
- MJO
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- March 13, 2018
- Terminated
- November 2, 2020
- Address
- 1800 Pyramid Pl, Memphis, TN, 38132-1703
Description
Medtronic Prestige LP(TM) Cervical Disc System with Streamlined Instruments, DRILL GUIDES: a) REF 6975205, 5mm b) REF 6975206, 6mm c) REF 6975207, 7mm Product Usage: The PRESTIGE LP reusable instruments are intended for the surgical implantation of the PRESTIGE LP Cervical Disc. Medtronic reusable instruments must be sterilized prior to initial use and must be cleaned and sterilized prior to each re-use.
The firm received complaints of drill bits breaking during use with the Prestige LP(TM) Cervical Disc System. Subsequent investigation demonstrated the interaction of the drill guide and the drill bit may lead to drill bit breakage.
Medtronic distributed a voluntary removal notice via FedEx courier service to impacted U.S. consignees on 13MARCH 2018. The notice requested the following: "Instructions to Customers: 1.) Please locate and remove the impacted product from normal storage locations. Do not use this product. 2.) Complete the enclosed Customer Confirmation Form and return via fax to 763-367-8134, or via e-mail to [email protected]. 3.) As applicable, contact your Medtronic Sales Representative to coordinate the return of affected product."
US Nationwide Distribution
231 units