FDA Recall Open, Classified

BEQ-HMO 50000-USA/ QUADROX-i Small, Oxygenator used during cardiopulmonary bypass Item no. 701067905

Recall: Z-1636-2023 · Initiated May 18, 2023

Recall

Recall Number
Z-1636-2023
Event Number
92003
Firm
Maquet Medical Systems USA
FEI Number
3008355164
Product Code
DTZ
Status
Open, Classified
Root Cause
Pending
Initiated
May 18, 2023
Posted
May 26, 2023
Address
45 Barbour Pond Dr, Wayne, NJ, 07470-2094

Description

BEQ-HMO 50000-USA/ QUADROX-i Small, Oxygenator used during cardiopulmonary bypass Item no. 701067905

Reason

1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory's sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas. Both issues may compromise the sterile barrier of the packaging, and exposure to non-sterile product may result in inflammation, infection, sepsis, and ischemia.

Action

On May 18, 2023, the firm notified customers of the recall via an Urgent Medical Device Removal letter. Actions to be taken by the customer: If an affected device is already in use, please use according to normal practices. Monitor the patient for signs and symptoms of inflammation, infection, sepsis, and ischemia. If detected, treat according to clinical protocols. Please remove any affected product in inventory from areas of use. Unopened and unexpired affected product is eligible for credit upon return. Product Return: Please contact Getinge Customer Service at 1-888-943-8872 option 2 between the hours of 6:00 AM and 5:00 PM Pacific Standard Time to request a return authorization (RMA) and shipping instructions to return any affected product. Distributors who have shipped affected products to customers should forward the notification to customers' attention for appropriate action. Maquet Cardiopulmonary GmbH is working with all possible urgency on the redesign and revalidation of a packaging solution, which is anticipated to be available by Q4 2023.

Distribution

Domestic distribution nationwide.

Quantity

1,784 (US)