FDA Recall Terminated

Dimension clinical chemistry system, Sirolimus (SIRO) Flex reagent cartridge (DF306 SMN 10464331) Lot EB6064, Open Well Instability The SIRO method is an in vitro diagnostic test for the quantitative measurement of sirolimus in whole blood on the Dimension clinical chemistry system.

Recall: Z-1633-2016 · Initiated February 9, 2016

Recall

Recall Number
Z-1633-2016
Event Number
73442
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2517506
Product Code
NRP
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
February 9, 2016
Terminated
June 12, 2018
Address
500 Gbc Dr PO BOX 6101, Ms 514, Newark, DE, 19702-2466

Description

Dimension clinical chemistry system, Sirolimus (SIRO) Flex reagent cartridge (DF306 SMN 10464331) Lot EB6064, Open Well Instability The SIRO method is an in vitro diagnostic test for the quantitative measurement of sirolimus in whole blood on the Dimension clinical chemistry system.

Reason

Siemens Healthcare Diagnostics has received customer complaints regarding QC shifts and imprecision after one day when using Dimension SIRO lot EB6064. They have confirmed SIRO lot # EB6064 does not meet the open well stability claim of 2 days.

Action

Siemens sent an "URGENT MEDICAL DEVICE RECALL" letter dated February 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Actions to be Taken by the Customer: Discontinue use of and discard the kit lot listed in Table 1. Please review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the product listed in Table 1, immediately contact your local Siemens Customer Care Center or your local Siemens technical support representative. Please retain this letter with your laboratory records, and forward this letter to those who may have received this product. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Customer Care Center or your local Siemens technical support representative. For further questions, please call (302) 631-0528.

Distribution

Worldwide distribution and US including Puerto Rico and to the states of : TX, FL, CA, LA, DC, GA, IL, AR, WI, MI and MD.

Quantity

394