FDA Recall Terminated

Stryker Howmedica Osteonics RESTORATION Acetabular Augment System; Howmedica Osteonics Corp., A subsidiary of Stryker Corporation Made in USA Sterile. Hip prosthesis component. The RESTORATION Acetabular Wedge Augments provide multiple options to address the wide range of bone deficiencies encountered in acetabular revision. The augments provide support for the shell in the acetabulum with superior and/or posterior defects.

Recall: Z-1633-2015 · Initiated August 15, 2014

Recall

Recall Number
Z-1633-2015
Event Number
70994
Firm
Stryker Howmedica Osteonics Corp.
FEI Number
3003070421
Product Code
LPH
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
August 15, 2014
Posted
May 19, 2015
Terminated
January 5, 2018
Address
325 Corporate Dr, Mahwah, NJ, 07430-2006

Description

Stryker Howmedica Osteonics RESTORATION Acetabular Augment System; Howmedica Osteonics Corp., A subsidiary of Stryker Corporation Made in USA Sterile. Hip prosthesis component. The RESTORATION Acetabular Wedge Augments provide multiple options to address the wide range of bone deficiencies encountered in acetabular revision. The augments provide support for the shell in the acetabulum with superior and/or posterior defects.

Reason

The peel strength of the inner blister may have been below internal validated requirements. The strength of the seal is directly related to long term shelf life of the sterile barrier.

Action

The firm, Stryker Orthopaedics, notified their Branches/Agencies via email on 8/15/2014 and sent an "Urgent Product Recall" letter dated 8/15/2014 with a Product Recall Acknowledgement Forms to their Branches/Agencies and consignees. The letter describes the product, issue, potential hazards and actions to be taken. The consignees were instructed to complete and return the attached Product Recall Acknowledgment Form within 5 days via fax to 855-251-3635; and return the affected product to the attention of Regulatory Compliance, Stryker Orthopaedics, 325 Corporate Drive, Mahwah NJ 07430. If you have any questions, feel free to contact Sr. Regulatory Compliance Specialist to 201-831-5826.

Distribution

Worldwide Distribution: US (nationwide) including PR and countries of: Australia, Canada, China, Brazil, France, Hong Kong, India, Italy, Mexico, Netherlands, Singapore, South Africa, Sweden, and Spain.

Quantity

1,147 in total