Atellica CH Toxicology Calibrator (TOX CAL)-The Atellica CH Toxicology Calibrator (TOX CAL) is for in vitro diagnostic use in calibrating the Acet, ETOH and Sal assays using the Atellica CH Analyzer Siemens Material Number (SMN): 11099440
Recall
- Recall Number
- Z-1632-2023
- Event Number
- 92179
- Firm
- Siemens Healthcare Diagnostics, Inc.
- FEI Number
- 2432235
- Product Code
- JLX
- Status
- Open, Classified
- Root Cause
- Under Investigation by firm
- Initiated
- April 5, 2023
- Posted
- May 25, 2023
- Address
- 511 Benedict Ave, Tarrytown, NY, 10591-5005
Description
Atellica CH Toxicology Calibrator (TOX CAL)-The Atellica CH Toxicology Calibrator (TOX CAL) is for in vitro diagnostic use in calibrating the Acet, ETOH and Sal assays using the Atellica CH Analyzer Siemens Material Number (SMN): 11099440
Reassignment of the Atellica CH Toxicology Calibrators (TOX CAL) and ADVIA Chemistry ToxAmmonia Calibrators for the Atellica CH Salicylate (Sal) Assay due to a positive bias
Siemens issued Urgent Medical Device Corrections (UMDC) ACHC23-03.A.US and ACHC23- 03.A.US.CHC to US customers via FedEx beginning on 04/05/2023. An Urgent Field Safety Notices (UFSN) ACHC23-03.A.OUS and ACHC23-03.A.OUS.CHC issued to OUS Siemens reporting sites with affected customers, for distribution and regional reporting to their local competent authorities beginning on 04/05/2023. Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Until the updated lot-specific value sheets are available on Document Library, keep a copy of this letter as a reference for the updated salicylate calibrator values. Perform the instructions provided in Additional Information section below. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
US Nationwide and Worldwide Distribution Foreign: Argentina Australia Austria¿¿¿¿¿¿¿¿ Bahrain¿¿¿¿¿¿¿¿ Bangladesh Belgium¿¿¿¿¿¿¿¿ Bosnia¿Herzeg.¿ Brazil Canada Chile China Colombia Czech¿Republic¿ Egypt¿¿¿¿¿¿¿¿¿¿ Estonia¿¿¿¿¿¿¿¿ Finland¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Georgia¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Hong Kong Hungary¿¿¿¿¿¿¿¿ India Israel¿¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Latvia¿¿¿¿¿¿¿¿¿ Malaysia Mexico Netherlands¿¿¿¿ Norway¿¿¿¿¿¿¿¿¿ Pakistan¿¿¿¿¿¿¿ Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Republic of Korea Saudi¿Arabia¿¿¿ Serbia¿¿¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ South Africa Spain¿¿¿¿¿¿¿¿¿¿ Sweden¿¿¿¿¿¿¿¿¿ Taiwan Turkey¿¿¿¿¿¿¿¿¿ United¿Kingdom¿ Uruguay Vietnam
6546 units US; 3681 units OUS