FDA Recall Terminated

natus newborn care Olympic Brainz Monitor, REF OBM00002. electroencephalograph

Recall: Z-1630-2020 · Initiated March 5, 2020

Recall

Recall Number
Z-1630-2020
Event Number
85189
Firm
Natus Neurology DBA Excel Tech., Ltd. (XLTEK)
FEI Number
3002809369
Product Code
OMA
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
March 5, 2020
Posted
April 1, 2020
Terminated
August 9, 2021
Address
2568 Bristol Cir, Oakville Canada

Description

natus newborn care Olympic Brainz Monitor, REF OBM00002. electroencephalograph

Reason

The OBM system functional test (impedance and noise) was not carried out on the affected items prior to release to the customer. The manufacturer cannot attest to the system functionality of the affected items.

Action

Natus initiated the recall by letter on 03/05/2020. the firm asked customers to return the affected units to Natus Medical Incorporated, 5900 First Avenue, Seattle, WA 98108, USA. Testing will be performed by Natus on the returned unit and will be returned to the consignee following completion of testing. Natus will provide a temporary replacement unit during this time.

Distribution

Worldwide distribution. US nationwide, Australia, Dominican Republic, United Arab Emirates, Israel, Italy, Hong Kong, Korea, Hungary, Switzerland, Colombia, and United Kingdom

Quantity

40 units