FDA Recall Terminated

Component for Laser Scanning Microscopes (LSM 700, LSM 800 and LSM 900)

Recall: Z-1629-2020 · Initiated October 2, 2019

Recall

Recall Number
Z-1629-2020
Event Number
84163
Firm
Zeiss, Carl Inc
FEI Number
2431026
Product Code
RIB
Status
Terminated
Root Cause
Radiation Control for Health and Safety Act
Initiated
October 2, 2019
Terminated
July 25, 2024
Address
1 Zeiss Dr, Thornwood, NY, 10594-1939

Description

Component for Laser Scanning Microscopes (LSM 700, LSM 800 and LSM 900)

Reason

When the user tilts the transmitted light arm of the microscope to the end of its axis range, laser safety can be bypassed and result in emission of class lIlb laser radiation. This issue arises when a ZEISS inverted stage stand (Axio Observer or Axio Vert) is used in as a component of the LSM 700, 800 and 900 system

Action

The Company will immediately notify by fax or direct contact (when available) and/or express courier each consignee purchaser and/or known user site with the laser safety notice which advises of the issue, the risk involved and how to minimize the risk of occurrence.

Distribution

Worldwide Distribution - US Nationwide

Quantity

US: 600; Global: 1600