FDA Recall Open, Classified

VITROS Chemistry Products HbA1c Reagent Kit- IVD for the quantitative determination of percent glycated hemoglobin A1c (DCCT/NGSP) and mmol/mol hemoglobin A1c (IFCC) in human whole blood. Product Code: 6842905

Recall: Z-1628-2023 · Initiated April 17, 2023

Recall

Recall Number
Z-1628-2023
Event Number
92254
Firm
Ortho-Clinical Diagnostics, Inc.
FEI Number
2250051
Product Code
PDJ
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
April 17, 2023
Posted
May 24, 2023
Address
1001 Us Highway 202, Raritan, NJ, 08869-1424

Description

VITROS Chemistry Products HbA1c Reagent Kit- IVD for the quantitative determination of percent glycated hemoglobin A1c (DCCT/NGSP) and mmol/mol hemoglobin A1c (IFCC) in human whole blood. Product Code: 6842905

Reason

Potential for both unexpected positive and negative HbA1c results, may result in misdiagnosis

Action

Ortho Clinical Diagnostics (Ortho) issued to Dealers and End-Users Urgent Product Correction Notification Potential Bias Observed in VITROS Chemistry Products HbA1c Reagent Kit on 4/17/23. Letter states reason for recall, health risk and action to take: In an abundance of caution, the required actions are applicable to all lots within expiry ( as well as future lots until your laboratory receives an additional notification from Ortho. However, while data is limited, a review of the release testing indicated no other lots were impacted. Reagent Pack Used for Calibration o If your system currently has multiple HbA1c reagent packs, leave one reagent pack on the system and unload all other reagent packs and store at 2-8 oC (36-46 oF) until needed. o Calibrate the one HbA1c reagent pack on your system. o Perform QC after successful calibration. QC must be acceptable and subsequently performed every 8 hours of testing, for the first 2 days that the reagent pack is onboard the system. o If QC fails with a negative bias, troubleshoot according to your laboratory s procedure. o If QC fails with a positive bias, load a new pack and re-calibrate following the required actions. Review the patient results since the last passing QC by your Medical Director. o After 2 days of passing within your laboratory s established QC range, performing QC every 8 hours is no longer necessary on that pack. Refer to Testing section below for QC requirements after calibration. Reagent Pack Loading Only have one reagent pack loaded onboard your VITROS System at a time. " Reagent Pack Used for Testing o Once the reagent pack used for calibration is depleted, QC must be performed upon loading each reagent pack and must be within the laboratory s established QC range prior to patient testing. o If the QC results are outside your laboratory s established range with a negative bias, then discard that reagent pack and load a new pack to perform QC again. o Ortho s observed data does not demonstrate

Distribution

US Nationwide- Worldwide Distribution. Foreign: Australia Belgium Brazil Bermuda HM12 Canada L3R 4G5 Chile China Colombia Denmark France France Germany India Italy Japan Mexico Norway Portugal Russia Singapore 609917 Spain Sweden The Netherlands United Kingdom

Quantity

23751