FDA Recall Terminated

-Scorpion Needle, For Use With Scorpion Suture Passer, (Blue). Ref. AR-13990N Lot 123263 -SureFire Scorpion Needle (purple). Ref. AR-13991N Lot 123263 and 122279 -Suture Passing Wire. Ref. AR-1255-18 Lot 123190 Manual surgical instrument for general use.

Recall: Z-1627-2013 · Initiated May 3, 2013

Recall

Recall Number
Z-1627-2013
Event Number
65205
Firm
Surgical Instrument Service And Savings, Inc.
FEI Number
3002299608
Product Code
GDF
Status
Terminated
Root Cause
Packaging process control
Initiated
May 3, 2013
Posted
July 1, 2013
Terminated
July 11, 2013
Address
2747 Sw 6th St, Redmond, OR, 97756-7109

Description

-Scorpion Needle, For Use With Scorpion Suture Passer, (Blue). Ref. AR-13990N Lot 123263 -SureFire Scorpion Needle (purple). Ref. AR-13991N Lot 123263 and 122279 -Suture Passing Wire. Ref. AR-1255-18 Lot 123190 Manual surgical instrument for general use.

Reason

The seal which maintains a sterile barrier for reprocessed medical devices became compromised to the point where product may fall out of the pouch.

Action

MEDISISS sent a Correction Immediate Action Request letter dated May 2, 2013, to all affected customes. The letter identified the product, the problem, and the action to be taken by the customer. The letter directed the customer to inspect the products prior to use for seal integrity as specified on the product labeling: "Sterile unless package has been damaged or opened. Do not use if package is damaged." This will be followed up by a visit from the customer's sales representatives who are tasked with inspecting all the lots that may be remaining in the customer's inventory. The sales representative is directed to complete a "Verification of Seal Inspection" form and return it to MEDISISS Customer Relations. Devices with unacceptable seals are to be returned to MEDISISS. The sales representative is responsible for obtaining a shipping label from customer service to enable the devices to be returned to the recalling firm via Fed Ex. Customers with questions were instructed to call 866-866-7477.

Distribution

Nationwide Distribution.

Quantity

6 units