FDA Recall Terminated

Gamma 3 System Set Screwdriver, Flexible Shaft; 4mm, This product is non-sterile Stryker Trauma GmbH, Germany; Distributed by Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430 - Orthopedic Device. The Gamma3 Locking Nail Systems are intramedullary short and long fixation nails intended for control of rotation and shortening of fractured femurs. The screwdriver is used to place the set screw in the nail.

Recall: Z-1626-2008 · Initiated February 22, 2008

Recall

Recall Number
Z-1626-2008
Event Number
46963
Firm
Stryker Howmedica Osteonics Corp.
FEI Number
3003070421
Product Code
HXX
Status
Terminated
Root Cause
Finished device change control
Initiated
February 22, 2008
Posted
September 16, 2008
Terminated
October 25, 2010
Address
325 Corporate Dr, Mahwah, NJ, 07430-2002

Description

Gamma 3 System Set Screwdriver, Flexible Shaft; 4mm, This product is non-sterile Stryker Trauma GmbH, Germany; Distributed by Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430 - Orthopedic Device. The Gamma3 Locking Nail Systems are intramedullary short and long fixation nails intended for control of rotation and shortening of fractured femurs. The screwdriver is used to place the set screw in the nail.

Reason

Stryker Trauma has become aware that five lots of the Gamma3 Set Screwdriver have hexagon tips that were not hardened to specification. If the screwdriver tip were to deform intra-operatively, the correct positioning of the set screw would not be possible, which could lead to lateral or medial migration of the lag-screw.

Action

On February 22, 2008 Urgent Product Recall notification letters were sent to Stryker Branches/Agencies, Risk Management Departments at hospitals and surgeons via Federal Express. If you have any questions contact Rita Intorrella at (201) 831-5825 or the Hip Fracture Team at (201) 831-5291.

Distribution

Class II Recall - Nationwide Distribution.

Quantity

43 total (single and kits)