FDA Recall Terminated

Deltec Cozmo Insulin Pump, Model 1800 Smiths Medical International Ltd WD24 4LG UK, Smiths Medical MD, Inc., St Paul, MN 55112. This Pump is similar to the currently distributed Deltec Cosmo Insulin Pump Model 1700 pump, with the exception of updated software that enhanced certain features and added a food data base to allow users to better manage dietary requirements in conjunction with the use of their insulin pump. REF 21-1801-81, 21-1803-81, 21-1805-49/51/81, 21-1806-51/81, 21-1807-49

Recall: Z-1623-2008 · Initiated March 10, 2008

Recall

Recall Number
Z-1623-2008
Event Number
47753
Firm
Smiths Medical MD, Inc.
FEI Number
2183502
Product Code
LZG
Status
Terminated
Root Cause
Component change control
Initiated
March 10, 2008
Posted
August 15, 2008
Terminated
January 13, 2011
Address
1265 Grey Fox Rd, Saint Paul, MN, 55112-6929

Description

Deltec Cozmo Insulin Pump, Model 1800 Smiths Medical International Ltd WD24 4LG UK, Smiths Medical MD, Inc., St Paul, MN 55112. This Pump is similar to the currently distributed Deltec Cosmo Insulin Pump Model 1700 pump, with the exception of updated software that enhanced certain features and added a food data base to allow users to better manage dietary requirements in conjunction with the use of their insulin pump. REF 21-1801-81, 21-1803-81, 21-1805-49/51/81, 21-1806-51/81, 21-1807-49

Reason

Smith Medical discovered an issue with a motor component in specific serial numbers of the Deltec Cozmo Insulin Pump, Model Numbers 1700 and 1800. This issue affects how the motor operates and may cause an over-delivery of insulin which could result in injury to the user.

Action

Consignees were mailed an Urgent Medical Device Recall letter dated March 10, 2008. The Urgent letter addressed to "Distributors, Clinicans, Patients and Other Customers" contains Description of the Problem, Results of Investigation, and Advise on Action to be taken by the User. The patient users will be required to return a Recall Confrirmation Form. The USA distributors, USA prescribing physicians and the USA managed Care Accounts (Insurance Companies) will be mailed also a cover letter.

Distribution

Class II Recall - Worldwide Distribution --- USA including states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME. RO, MN, MO, MS, ME, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of Greece, Netherland, Sweden, France, Germany, Turkey, China, Great Britain, Israel, Italy, Kuwait, New Zealand, Australia.

Quantity

1066