FDA Recall Open, Classified

Aligned Medical Solutions Minor Hand Pack, REF: AMS6539E, containing P 2 Cover Light Handle Blue STE, Sterile EO

Recall: Z-1620-2023 · Initiated April 24, 2023

Recall

Recall Number
Z-1620-2023
Event Number
92288
Firm
Windstone Medical Packaging, Inc.
FEI Number
1000125955
Product Code
KDD
Status
Open, Classified
Root Cause
Unknown/Undetermined by firm
Initiated
April 24, 2023
Posted
May 23, 2023
Address
1602 4th Ave N, Billings, MT, 59101-1521

Description

Aligned Medical Solutions Minor Hand Pack, REF: AMS6539E, containing P 2 Cover Light Handle Blue STE, Sterile EO

Reason

Light handle covers may separate from the light handle (detach and fall off) during use, this could result in a procedure delay and/or impact the sterile field which could result in an injury to the patient

Action

On April 24, 2023, Aligned Medical Solutions issued a "Urgent: Medical Device Recall" notification via E-Mail. Customers were asked to take the following actions: 1. Immediately check your inventory for the voluntary recalled products listed above and place them under quarantine. 2. Due to the manufacturing process any finished sterilized packs remaining in inventory, or at your facility will require over labeling for the all products affected by this recall. Packs will be labeled with a sticker attached to each pack. 3. Complete the attached Recall Reply Form listing all inventory of affected product by lot number and quantity 4. If you have further distributed this products, please identify your distribution partners, and notify them at once of the product recall. Your notification to your customers should include a copy of this notification letter. 5. Please complete the Recall Reply Form within 5 business days and return via email to [email protected] even if you have no affected product on hand. 6. Indicate on the Recall Reply Form if a replacement STERIS product is needed. 7. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

Distribution

US: MA OUS: None

Quantity

90 kits