FDA Recall Terminated

GLU Reagent (GLU cartridge assay) used on Synchron LX and UniCel DxC Synchron Clinical Systems, Part Number 442640; SYNCHRON and UniCel Systems and Part Numbers: Synchron LX20, Part Number: 466200, Synchron LX20 PRO, Part Number: 476100, Synchron LXI 725, Part Number: 476501, UniCel DxC 600, Part Number: A10405, UniCel 600 PRO, Part Number: A11810, UniCel DxC 600i, Part Number: A27318, UniCel DxC 800, Part Number: A11816, UniCel DxC 800 PRO, Part Number: A11812, UniCel DxC 880i, Part Number: A59102, UniCel DxC 660i, Part Number: A64871, UniCel DxC 680i, Part Number: A64903, UniCel DxC860i, Part Number: A64935 (1) The Synchron LX and UniCel DxC Synchron Clinical Systems are fully automated, computer controlled clinical chemistry analyzers designed for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood (Sample type is chemistry dependent). (2) GLUCm reagent, when used in conjunction with SYNCHRON LX System(s), UniCel DxC 600/800 System(s) and SYNCHRON Systems AQUA CAL 1 and 2, is intended for the quantitative determination of Glucose concentration in human serum, plasma, urine or cerebrospinal fluid (CSF).

Recall: Z-1619-2011 · Initiated August 10, 2010

Recall

Recall Number
Z-1619-2011
Event Number
56746
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
CEM
Status
Terminated
Root Cause
Other
Initiated
August 10, 2010
Posted
March 9, 2011
Terminated
October 17, 2013
Address
200 S Kraemer Blvd, Brea, CA, 92821-6208

Description

GLU Reagent (GLU cartridge assay) used on Synchron LX and UniCel DxC Synchron Clinical Systems, Part Number 442640; SYNCHRON and UniCel Systems and Part Numbers: Synchron LX20, Part Number: 466200, Synchron LX20 PRO, Part Number: 476100, Synchron LXI 725, Part Number: 476501, UniCel DxC 600, Part Number: A10405, UniCel 600 PRO, Part Number: A11810, UniCel DxC 600i, Part Number: A27318, UniCel DxC 800, Part Number: A11816, UniCel DxC 800 PRO, Part Number: A11812, UniCel DxC 880i, Part Number: A59102, UniCel DxC 660i, Part Number: A64871, UniCel DxC 680i, Part Number: A64903, UniCel DxC860i, Part Number: A64935 (1) The Synchron LX and UniCel DxC Synchron Clinical Systems are fully automated, computer controlled clinical chemistry analyzers designed for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood (Sample type is chemistry dependent). (2) GLUCm reagent, when used in conjunction with SYNCHRON LX System(s), UniCel DxC 600/800 System(s) and SYNCHRON Systems AQUA CAL 1 and 2, is intended for the quantitative determination of Glucose concentration in human serum, plasma, urine or cerebrospinal fluid (CSF).

Reason

The recall was initiated because Cartridge Glucose (GLU) on UniCel DxC and Synchron LX instruments may give falsely high results for moderately hemolysed samples. The Impact: (1) Neonatal samples drawn by lancets are more likely to be hemolysed to a degree sufficient to cause interference with glucose estimation and are a special at-risk population, (2) Other hemolysed samples may be similar

Action

The firm, Beckman Coulter, sent an "URGENT: PRODUCT CORRECTIVE ACTION" letter dated August 10, 2010 with an enclosed fax-back Customer Response Form to the customers. The letter provided the customers with an explanation of the problem identified, the impact and an action to be taken. The customers were instructed to run samples using another method/technology and to complete and return the enclosed FAX BACK RESPONSE FORM via fax to (714) 961-4232 or mail to: Beckman Coulter, Inc., Attn: Regulatory Affairs, 250 S. Kraemer Blvd., PO Box 8000, Brea, CA 92822. Firms with questions regarding this Product Corrective Action letter, were instructed to call our Customer Support Center (Hotline) at (800) 854-3633 in the United States and Canada http://www.beckmancoulter.com/customersupport/support/default.asp.

Distribution

Worldwide distribution: USA and country: Canada.

Quantity

81,871 units