FDA Recall Terminated

EXCOR Pediatric Ventricular Assist Device Blood Pumps, Models: P10P-001 (10 ml blood pump); P15P-001 (10 ml blood pump); P25P-001x01 (25 ml blood pump); P30P-001x01 (30 ml blood pump), P50P-001 (50 ml blood pump); P60P-001 (60 ml blood pump). Recall occurred in 2018, and was retrospectively reported.

Recall: Z-1618-2023 · Initiated May 18, 2018

Recall

Recall Number
Z-1618-2023
Event Number
92132
Firm
BERLIN HEART GMBH Wiesenweg 10 Berlin Germany
FEI Number
3004582654
Product Code
DSQ
Status
Terminated
Root Cause
Component design/selection
Initiated
May 18, 2018
Posted
May 25, 2023
Terminated
June 6, 2024

Description

EXCOR Pediatric Ventricular Assist Device Blood Pumps, Models: P10P-001 (10 ml blood pump); P15P-001 (10 ml blood pump); P25P-001x01 (25 ml blood pump); P30P-001x01 (30 ml blood pump), P50P-001 (50 ml blood pump); P60P-001 (60 ml blood pump). Recall occurred in 2018, and was retrospectively reported.

Reason

There is a potential for a disruption in one or more of the triple layer membranes of the ventricular assist device. Recall occurred in 2018, and was retrospectively reported.

Action

On May 18, 2018, Berlin Heart issued a Safety Communication via E-Mail to affected consignee. The Safety Communication asked consignees to take the following actions: 1. Always follow the information and instructions provided in the Instructions for Use that have been provided to you by Berlin Heart. Information related to the specific issues discussed within the letter is included in Chapter 12 starting on page 115, specifically Section 12.5, and 12.51. 2. Additional information on troubleshooting is found in Chapter 16 starting on page 161. 3. If you have any questions or concerns regarding the communication, please contact Clinical Affairs team directly at 281-863-9700

Distribution

Distribution to US states of MA, TX, WI, CA, SC, FL, CO, NY, PA, NC, UT, GA, MN, OH, AZ, AL, and Canada

Quantity

335 units