FDA Recall Terminated

COULTER DxH Diluent, Catalog No. 628017, Part No. A59956. For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells.

Recall: Z-1616-2015 · Initiated April 13, 2015

Recall

Recall Number
Z-1616-2015
Event Number
71045
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
GIF
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
April 13, 2015
Posted
May 15, 2015
Terminated
October 1, 2015
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

COULTER DxH Diluent, Catalog No. 628017, Part No. A59956. For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells.

Reason

Beckman Coulter has received an increased number of complaints related to Hemoglobin (HGB) on UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems. The use of the DxH Diluent lots listed in the recall notice may eventually result in a compromised HGB chamber.

Action

An Urgent Medical Device Recall letter dated 4/13/15 was sent to customers who purchased the Coulter DxH Diluent. The letter informs customers that Beckman Coulter has received an increased number of complaints related to Hemoglobin (HGB) on UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems. The letter informs customers of the problems identified and the actions to be taken. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with questions are instructed to contact Customer Support Center via http://www.beckmancoulter.com/customersupport/support or call (800) 526-7694 in the US and Canada. Customers outside the US and Canada are instructed to contact their local Beckman Coulter Representative.

Distribution

Worldwide Distribution -- Argentina; Brazil; Brunei Darussalam; Canada; Chile; Colombia; Hong Kong; India; Lebanon; Macao; Mexico; Panama; Philippines; Qatar; Singapore; Taiwan; Thailand; Trinidad and Tobago; United States, including Puerto Rico; and Uruguay.

Quantity

829,888 units total (659,584 units in US)