Hitachi AIRIS, AIRIS II, AIRIS Elite MRI Scanner Compensation Coil. Part numbers: MR-FC-2EO, MR-FC-3EO

Recall: Z-1616-2010 · Initiated April 19, 2010

Recall

Recall Number
Z-1616-2010
Event Number
55555
Firm
Hitachi Medical Systems America Inc
FEI Number
1528028
Product Code
MOS
Status
Terminated
Root Cause
Other
Initiated
April 19, 2010
Posted
May 19, 2010
Terminated
September 21, 2012
Address
1959 Summit Commerce Park, Twinsburg, OH, 44087

Description

Hitachi AIRIS, AIRIS II, AIRIS Elite MRI Scanner Compensation Coil. Part numbers: MR-FC-2EO, MR-FC-3EO

Reason

Hitachi determined that the mounting method for a Compensation Coil did not include an appropriate factor of safety, so it is possible that the coil could fall, contacting the patient table and potentially putting a patient at risk.

Action

On 4/15/2010 the firm sent Device Correction Notice letters to their customers identifying the affected product and informing them of the potential problem. The letter stated that the firm will change the coil mounting method used at the customer's facility, if necessary, at no cost to the customer. A service rep will contact them to schedule a visit to correct the problem.

Distribution

Nationwide Distribution -- AL, AR, AZ, CA, CT, DC, DE, FL, ID, IL, IN, KY, LA, MD, MO, NJ, NY, OH, OK, OR, RI, TN, TX, VA & WA.

Quantity

56 devices