Hitachi AIRIS, AIRIS II, AIRIS Elite MRI Scanner Compensation Coil. Part numbers: MR-FC-2EO, MR-FC-3EO
Recall
- Recall Number
- Z-1616-2010
- Event Number
- 55555
- Firm
- Hitachi Medical Systems America Inc
- FEI Number
- 1528028
- Product Code
- MOS
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 19, 2010
- Posted
- May 19, 2010
- Terminated
- September 21, 2012
- Address
- 1959 Summit Commerce Park, Twinsburg, OH, 44087
Description
Hitachi AIRIS, AIRIS II, AIRIS Elite MRI Scanner Compensation Coil. Part numbers: MR-FC-2EO, MR-FC-3EO
Hitachi determined that the mounting method for a Compensation Coil did not include an appropriate factor of safety, so it is possible that the coil could fall, contacting the patient table and potentially putting a patient at risk.
On 4/15/2010 the firm sent Device Correction Notice letters to their customers identifying the affected product and informing them of the potential problem. The letter stated that the firm will change the coil mounting method used at the customer's facility, if necessary, at no cost to the customer. A service rep will contact them to schedule a visit to correct the problem.
Nationwide Distribution -- AL, AR, AZ, CA, CT, DC, DE, FL, ID, IL, IN, KY, LA, MD, MO, NJ, NY, OH, OK, OR, RI, TN, TX, VA & WA.
56 devices